TY - JOUR
T1 - Influence of in situ, sound-level calibration on distortion-product otoacoustic emission variability
AU - Scheperle, Rachel A.
AU - Neely, Stephen T.
AU - Kopun, Judy G.
AU - Gorga, Michael P.
PY - 2008
Y1 - 2008
N2 - Standing waves can cause errors during in-the-ear calibration of sound pressure level (SPL), affecting both stimulus magnitude and distortion-product otoacoustic emission (DPOAE) level. Sound intensity level (SIL) and forward pressure level (FPL) are two measurements theoretically unaffected by standing waves. SPL, SIL, and FPL in situ calibrations were compared by determining sensitivity of DPOAE level to probe-insertion depth (deep and "shallow") for a range of stimulus frequencies (1-8 kHz) and levels (20-60 dB). Probe-insertion depth was manipulated with the intent to shift the frequencies with standing-wave minima at the emission probe, introducing variability during SPL calibration. The absolute difference in DPOAE level between insertions was evaluated after correcting for an incidental change caused by the effect of ear-canal impedance on the emission traveling from the cochlea. A three-way analysis of variance found significant main effects for stimulus level, stimulus frequency, and calibration method, as well as significant interactions involving calibration method. All calibration methods exhibited changes in DPOAE level due to the insertion depth, especially above 4 kHz. However, SPL demonstrated the greatest changes across all stimulus levels for frequencies above 2 kHz, suggesting that SIL and FPL provide more consistent measurements of DPOAEs for frequencies susceptible to standing-wave calibration errors.
AB - Standing waves can cause errors during in-the-ear calibration of sound pressure level (SPL), affecting both stimulus magnitude and distortion-product otoacoustic emission (DPOAE) level. Sound intensity level (SIL) and forward pressure level (FPL) are two measurements theoretically unaffected by standing waves. SPL, SIL, and FPL in situ calibrations were compared by determining sensitivity of DPOAE level to probe-insertion depth (deep and "shallow") for a range of stimulus frequencies (1-8 kHz) and levels (20-60 dB). Probe-insertion depth was manipulated with the intent to shift the frequencies with standing-wave minima at the emission probe, introducing variability during SPL calibration. The absolute difference in DPOAE level between insertions was evaluated after correcting for an incidental change caused by the effect of ear-canal impedance on the emission traveling from the cochlea. A three-way analysis of variance found significant main effects for stimulus level, stimulus frequency, and calibration method, as well as significant interactions involving calibration method. All calibration methods exhibited changes in DPOAE level due to the insertion depth, especially above 4 kHz. However, SPL demonstrated the greatest changes across all stimulus levels for frequencies above 2 kHz, suggesting that SIL and FPL provide more consistent measurements of DPOAEs for frequencies susceptible to standing-wave calibration errors.
UR - http://www.scopus.com/inward/record.url?scp=47649086674&partnerID=8YFLogxK
U2 - 10.1121/1.2931953
DO - 10.1121/1.2931953
M3 - Article
C2 - 18646977
AN - SCOPUS:47649086674
SN - 0001-4966
VL - 124
SP - 288
EP - 300
JO - Journal of the Acoustical Society of America
JF - Journal of the Acoustical Society of America
IS - 1
ER -